Windows XP PC: C:\Documents and Settings\\Local Settings\Application Data\ Agilent_Technologies\CHEMMAIN.EXE_Url_xlvsr1c1b5wijktns4g1gcbrae3dl030\2.0.0.0\nfig.Windows 7/ Windows 10 PC: c:\users\\appdata\local\agilent_technologies\ chemmain.exe_url_xyz\2.0.0.0\nfig.Rename or delete the nfig file it can be found at the following location: Close all OpenLab ChemStation software app.ģ. Select the View tab and, in Advanced settings, select Show hidden files, folders, and drives and OK.Ģ.Select View > Options > Change folder and search options.Hidden items will need to be enabled from the File Explorer View menu in order to see appdata: It will be recreated automatically during the next ChemStation launch process.ġ. This error message is usually the result of a corruption in the nfig file or a corrupted instrument configuration (typically stemming from an unexpected shutdown or power outage). ChemStation will shutdown now", this article will provide you some basic steps in troubleshooting the issue. When your Agilent OpenLab ChemStation Edition software fails to launch with errors "Login failed with the following message: Configuration system failed to initialize.